EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Air-Shields TI-100 Transport Incubator

Air-Shields TI-100 Transport Incubator

Brand: Airshields
Web ID: 217
Condition: Refurbished

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

THIS ITEM IS CURRENTLY UNAVAILABLE.  However for hospital use we do have newer refurbished options so please reach out to our team.  We also carry infant incubators dedicated for educational and simulation purposes only.

---

The Air-Shields TI-100 Transport Incubator is available refurbished from Avante Health Solutions.

FEATURES

  • Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
  • It provides a means to control air temperature, oxygen concentrations, and relative humidity.
  • A double wall hood permits full visibility and provides an effective thermal barrier from the environment.

 

SPECIFICATIONS

Dimensions (with stand):

  • Length: 40.25 in. (102 cm)
  • Width: 22.25 in. (56.5 cm)
  • Height: 46.75 in. (118.7 cm) maximum, 34.75 in. (88.3 cm) minimum

Power Requirements:

  • AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
  • DC: 12 VDC, 115 Watts

Internal Battery:

  • Type: gel type, sealed and rechargeable.
  • Voltage: 12 VDC nominal
  • Quantity: 2
  • Charging Time: two batteries take 20 minutes to recharge from fully discharged.
  • Life expectancy: minimum 200 full charge/discharge cycles.

Chassis Leakage Current: 100 µA or less

Alarms:

  • High Temp: Actuates at displayed incubator temperature of 39° ± 0.5 °C.
  • Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
  • Power Failure: Actuates if main AC power fails and no DC back-up power is present.
  • Heater Temp: Actuates if heater temperature exceeds 77 °C nominal.
  • Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.

Temperature:

  • Set Point Range: 21.5 °C ± 1.0 °C to 38 °C
  • Rise Time: 30 minutes
  • Variability: 0.2 °C
  • Overshoot: 1.5 °C maximum
  • Uniformity: 1 °C
  • Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1 °C
  • Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5 °C
  • Observation Light Output: 10-foot candles at mattress level