BCI Mini-Torr Plus - NIBP / Oximeter Monitor with Printer

BCI Mini-Torr Plus - NIBP / Oximeter Monitor with Printer

Brand: Smiths BCI
Web ID: 410

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The Mini-TorR Plus Non-Invasive Blood Pressure Monitor provides a high quality reliable platform for your NIBP monitoring needs. Its combination of reliability and flexibility makes the Mini-Torr Plus® a economical starting point for your long-range NIBP monitoring needs.

Specifications

Physical Weight & Dimensions

  • Base Unit 3.24 (H) × 8.5 (W) × 5.5 (D) inches
  • With Temp 3.24 (H) × 10 (W) × 5.5 (D) inches
  • Base Unit 3.5 lbs (1.6 kg)
  • With Temperature 4.2 lbs (1.9 kg)

Display

  • Pulse Rate 3-digit LED display, 0.30 inches (7.62 mm) high
  • Pulse Strength Logarithmically scaled 10-segment
  • LED bar graph
  • NIBP 3-digit LED display
  • Systolic & Diastolic Pressures
  • 0.40 inches (10.16 mm) high
  • Mean Arterial Pressure
  • 0.30 inches (7.62 mm) high
  • Temperature 3-digit LED display, 0.40 inches (10.16 mm) high

NIBP

  • Adult: 20 to 250 mmHg
  • Neonate: 20 to 135 mmHg
  • Cuff Pressure ±3 mmHg or 2%
  • Determination 30 to 50 seconds typical
  • Time (max. 120 seconds in adult mode, 90 seconds in neonate mode)
  • Interval Times 1, 2, 2.5, 5, 10, 15, 20, 30, 45, 60, 90 mins
  • Calibration Factory calibrated
  • Inflation Pressure Adult: 50 to 300 mmHg
  • Neonate: 50 to 150 mmHg
  • Default Inflation Adult: 175 mmHg
  • Pressure Neonate: 100 mmHg
  • Alarm Range
  • Adult High: 15 to 285 mmHg & OFF
  • Low: 10 to 280 mmHg & OFF
  • Neonate High: 15 to 135 mmHg & OFF , Low: 10 to 130 mmHg & OFF
  • Alarm Default
  • Adult Systolic: High - 200 mmHg , Low - 90 mmHg
  • Diastolic: High - 105 mmHg, Low - 40 mmHg
  • Mean Arterial: High - 140 mmHg, Low - 50 mmHg

SpO2 (Optional)

  • Range 0 to 100 SpO2 (functional)
  • Accuracy ±2 counts at 70 to 100%
  • Averaging 8 beats
  • Pulse Tone Volume adjustable or OFF
  • Alarm Default Adult: High: 150 bpm, Low: 45 bpm
  • Neonate: High: 150 bpm, Low: 50 bpm

Temperature (Optional)

  • Range 84°F to 108°F (28.9°C to 42.2°C)
  • Accuracy ±2°F, ASTM E1112-86 (±0.1°C)
  • ±3 counts at 50 to 69%
  • Calibration Factory calibrated

Battery Life

Fully charged continuous-life 6 hours approximate, charge time 4 hours

Printer Output

  • RS232C
  • Data Format: 19200 baud

Integrated Printer (Optional)

Text only, no graphics. Patient data log or trend tables.

Regulatory Compliance

  • 510(k): K983796, K984618, K970801
  • ISO 9001 / EN46001