Monitor de Signos Vitales DRE Signal SP

Monitor de Signos Vitales DRE Signal SP

Marca: Avante
Web ID: 1189
Condición: Retired

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Ultra-lightweight and portable with up to 6 hours of battery life.

Measures Non-Invasive Blood Pressure, SpO2, Stat-Temperature.

SureTemp™ Thermometer provides an accurate reading within seconds

  • Suitable for adult, pediatric and neonatal patients
  • Nellcor SpO2 technology, and motion tolerant SPO2 and NIPB
  • Adjustable audible and visual alarms
  • Removable and rechargeable battery provides up to 6-hour working time
  • Integrated handle, full mounting solutions and integrated printer

Specifications

Physical Weight & Dimensions

  • Height: 240 mm
  • Width: 170 mm
  • Depth: 170 mm
  • Weight: < 3.5g (including a battery)

Safety

Meets the requirements of IEC60601 series, CE marking according to MDD93/42/EEC

Operation Environment

  • AC Power Req: 100-240VAC (± 10%), SO/60 Hz (± 3 Hz), 70VA
  • Temperature: 0-40°C (SureTemp™: 10-40°C)
  • Humidity: 15%-95%, non-condensing

Patient Range

Adult, pediatric and neonatal

NIBP

  • Method: Automatic oscillometric
  • Operation modes: Manual / Automatic / Continual
  • Auto measure time: Adjustable
  • Measurement unit: mmHg / kPa selectable
  • Measurement types: Systolic, Diastolic, Mean, Pulse Rate
  • Measurement range: Adult: 10-270mmHg; Pediatric: 10-200mmHg; Neonate: 10-135mmHg

Over-pressure

  • Protection: YES
  • Resolution: 1mmHg
  • Alarm: Systolic, Diastolic, Mean, Pulse Rate
  • Pulse Rate Range: 40-240 bpm
  • Pulse Rate Resolution: 1bpm
  • Pulse Rate Accuracy: ± 2bpm

Pulse Rate

Range and accuracy: 20-250 bpm ± 3 bpm, 251-300 bpm unspecified

Typical Measurement

Time: < 12s

SpO2

  • Measurement range: 0-100%
  • Resolution: 1%
  • Accuracy:
    • Adult/Pediatric: ±2% (70-100%)
    • Neonatal: ±3% (70-100%)
    • 0-69% unspecified
  • Alarm range: 0-100%
  • Pulse rate:
    • Range: 20-254bpm
    • Resolution: 1bpm
    • Accuracy: ± 3bpm
    • Alarm range: 20-254bpm
    • Storage Humidity: 10%RH to 95%RH

Nellcor OxiMax SpO2

  • Measurement range: 0-100%
  • Compatible sensors:
    • Nellcor sensor type: OxiMax series MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I, MAX-FAST, MAX-R,OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I, D-YS, D-YSE, D-YSPD, DS-100A, OXI-AIN and OXI-P/I
    • SpO2 accuracy:
      • 70%-100% ±2%; 0-69% unspecified (sensors: MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST)
      • 70%-100% ± 2.S%; 0-69% unspecified (sensors: OxiCliq A, OxiCliq N, OxiCliq P and OxiCliq I)
      • 70%-100% ± 3%; 0-69% unspecified (sensors: D-YS, DS-100A, OXI-A/N and OXI-P/I)
      • 0%-100% ±3.5%; 0-69% unspecified

SureTemp™ Thermometer

  • Measurement range: 25 - 44°C (77-111.2°F)
  • Resolution: In MONITOR mode: 0.1°C (0.2°F)
  • Accuracy:
    • In MONITOR mode: 25-32°C (77-89.6 °F) : ±0 .2°C (±0.3 °F) including 32°C (89.6 °F)
    • In MONITOR mode: 32~44°C (89.6-111.2 °F) : ±O.l °C (±0.2 °F) excluding 32°C (89.6 °F)