; Avante Waveline EZ Monitor Portátil

EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Avante Waveline EZ Monitor Portátil

Avante Waveline EZ Monitor Portátil

Web ID: 1230
Número de producto: 66012B2x (x) = additional suffix for varying configurations

0 Artículo está en su carrito Ver carrito ahora

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Waveline Ez

Monitor portátil con una pantalla táctil que es fácil de utilizar!

Viene con: ECG • Respiración • SpO2 • Presión Non-Invasiva • Temperatura.

Disponible con EtCO2 e imprimadora.

 

Este monitor destaca las últimas innovaciones y una artesanía impecable. El Avante Waveline EZ es la selección perfecta para los profesionales del sector médico quienes demandan precisión, funcionalidad y asequibilidad. Su pantalla táctil le ayuda evaluar la condición de sus pacientes rápidamente y con precisión, la cual resulta unamejor atención dada a sus pacientes.

 

Caracteristicas:

  • Pesa 6 lbs. Uno de los monitores multiparametros mas móviles en el mercado
  • Pantalla táctil provee al utilizador una disponibilidad inmediata
  • Puede medir simultáneamente el ECG de sensores múltiples
  • Detección avanzada de arrhythmia y ST
  • Capabilidad de mostrar tendencias en formas gráficas y tabulares
  • Se puede ver tres ondas a la vez en la pantalla
  • Apoyo respaldo de batería
  • Alarmas visuales y audibles
  • Memoria rápida de todas las medidas de presión

 

 

Specifications

Physical Dimensions

  • Height: ~8-1/5 inches
  • Width: 9 inches
  • Depth: 4-7/10 inches

Safety Approval & Quality System

  • Designed to meet IEC60601-1-1988, EN60601-1-1, EN60601-2
  • Class II Equipment, double insulated
  • Type BF applied parts
  • ISO9001 & EN46001 Certified

Power Requirements

  • Power Supply AC 90-264V/47-63Hz
  • Input Power ≤55VA
  • Fuses Two fuse sockets in the rear panel indicated by "FUSE", OI 5X20, 2A/250V
  • Battery 12V/4.0AH sealed lead-acid
  • Charge time ≥4 hours
  • Operating time ≥2 hours (full recharge)
  • Battery Charging Method Automatic
  • charging after monitor is connected to AC power supply (with charge protection function)
  • Discharge Protection When powered by battery, the monitor will be automatically turned off when battery power is almost used up

ECG

  • Input -3-lead ECG cable and standard AAMI line for connection
  • Input - (Optional) 5-lead ECG cable and standard AAMI line for connection
  • Patient Safety Standard IEC60601-1-1988
  • CMRR ≥60dB (Common Mode Rejection Ratio)
  • Heart Rate Range 20 ~ 254bpm ± 1bpm
  • Heart Rate Averaging 8 second average
  • ST Segment Range -0.8 ~ + 0.8mV
  • Interface AAMI 6-pin
  • Lead Selection I, II, III (3 lead mode)
  • I, II, III, aVR, aVL, aVF, V (5 lead mode) (ST and Arrhythmia analysis)
  • Lead Fault Alarm Audible, Visual Input 5-lead ECG patient cable
  • QRS Indicator Audible and Visual Alert
  • Waveform Storage 6 minutes
  • Sweep Speed 12.5/25/50 mm/sec
  • Gain Selection 4mV, 2mV, 1mV, 0.5mV, 0.25mV,Auto
  • Trends 2 hours -> 4 hours -> 8 hours -> 24 hours -> 48 hours
  • Patient Isolation Breakdown voltage 4000VAC 50Hz 60 seconds Leakage current <10μA
  • Frequency width Monitoring mode 0.5 ~ 40Hz (+0.4dB,-3.0dB) Surgery mode 0.5 ~ 20Hz (+0.4dB, -3.0dB) not calibration significant
  • Patient Drive Current <10μA
  • Enclosure Leakage Current <0.1mA
  • Maximum T Wave Rejection Capability 1.2mV
  • Heart Rate Alarm Response Time < 7 seconds
  • Aspect Ratio 0.24 ~ 0.6 sec/mV
  • Alarm Frequency Low alarm: 2-2.4kHz High alarm: 3-3.4kHz
  • Defibrillator Protected & ESIS Protected Tested with 5kV
  • Recovery Time Following Defibrillation <5 seconds

Respiration

  • Measurement Method Thoracic Impedance
  • Respiration Rate Range 0 ~ 100±1rpm
  • Accuracy ±2 rpm

Masimo SET SpO2

  • SpO2 Accuracy (non-motion):
  • Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
  • SpO2 Accuracy (motion):
  • Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
  • SpO2: ±2 %
  • PR: ±3 bpm
  • Modes:
  • Averaging mode: 2,4,8,10,12,14 and 16 s
  • Sensitivity: Normal, APOD and Maximum
  • PR Accuracy (non-motion):
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
  • PR Accuracy (motion):
  • Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
  • Measuring Range:
  • SpO2: 1~100 %
  • PR: 25~240 bpm
  • Perfusion: 0.02~20 %

Non-Invasive Blood Pressure (NIBP)

  • Method Automatic oscillometric
  • Parameters Systolic, diastolic, mean arterial pressure, pulse
  • Scale mmHg or kPa
  • Operating Modes Manual, Automatic, Continuous
  • Repeat Cycles 1 ~ 10, 15, 30, 60, 90, 120 minutes
  • Determination Systolic, Adult/pediatric 40 ~ 250mmHg (5.3 ~ 33.3kPa) Systolic, Neonate 20 ~ 160mmHg (2.7 ~ 21.3kPa) Diastolic, Adult/pediatric 10 ~ 180mmHg (1.3 ~ 24.0kPa) Diastolic, Neonate 10 ~ 140mmHg (1.3 ~ 18.7kPa)
  • Cuff Pressure Range Adult/pediatric 0 ~ 300mmHg (0 ~ 40.0kPa) Neonate 0 ~ 140mmHg (0 ~ 18.7kPa)
  • Initial Cuff Inflation Adult/pediatric 170±10mmHg (22.7±1.3kPa) Neonate 100±10mmHg (16.0±1.3kPa)
  • Deflation Pressure 30mmHg(4.0kPa) higher than the last systolic pressure
  • Cuff Inflation Rate No greater than 50mmHg/sec
  • Measurement Time Typical 25 seconds, Maximum 40 seconds, Typical Stat 20 seconds
  • Pressure Display Accuracy ±3mmHg
  • BP Pulse Rate Accuracy ±2% @ 40 ~240bpm
  • Cuff Neonate, infant, pediatric,standard adult

Temperature (Dual Channel)

  • Range 0 ~ 50°C
  • Probe YSI ® 400 Skin surface or rectal /esophageal
  • Scale Celsius
  • Accuracy ±0.1°C
  • Resolution 0.1°C

CO2

  • Type Side stream, mom-dispersive IR
  • CO2 Range 0-99mmHg
  • Scale mmHg/kPa
  • Accuracy ± 2mmHg (0-40mmHg) ± 5mmHg (41-76mmHg) ±10mmHg (77-99mmHg)
  • Calibration Automatic
  • Respiration Range 0-150rpm, ± 2rpm

TFT Color Display

  • Size 8 inches
  • Matrix 640 (H) × 480 (V) pixels

Recorder (Optional)

  • Type Built-in 2-channel thermal array recorder
  • Print mode Text or waveform
  • Waveforms Real time or alarm-triggered
  • Resolution 400dpi vertical, 800dpi horizontal
  • Annotations Time, date, vital sign readings