EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

DRE Trax Monitor De Signos Vitales

DRE Trax Monitor De Signos Vitales

Marca: Avante
Web ID: 1601
Condición: Retired
Número de producto: 66311x (x = additional suffix for varying configurations

0 Artículo está en su carrito Ver carrito ahora

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

El monitor DRE Trax es muy vers

Specifications

General

  • Display: 5.0 inch (Diagonal) color TFT
  • Resolution: 800 × 3(RGB) × 480
  • Trace: 2 waveforms
  • Waveforms: PLETH, ETCO2
  • Trend time: From 1 to 72 hours
  • Indicator:
    • Alarm Indicator
    • Power indicator
    • Pulse beep and alarm sound

NIBP

  • Measuring Technology: Automatic oscillating measurement
  • Cuff Inflating: <30s (0 ~ 300 mmH, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto, STAT
  • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
  • Pulse Rate Range: 30 bpm ~ 250 bpm
  • Measuring Range: Adult
  • SYS: 40 ~ 250 (mmHg)
  • DIA: 15 ~ 200 (mmHg)
  • Resolution: 1mmHg
  • Pressure Accuracy: Maximum Mean error: ±5mmHg
  • Maximum Standard deviation: 8mmHg
  • Overpressure Protection: Adult Mode: 280(mmHg)
  • Alarm Limit: SYS: 50 ~ 240 mmHg
  • DIA: 15 ~ 180 mmHg
  • Standards: Meets performance standards of ANSI/AAMI SP10:2002

SPO2

  • ASpO2: Anti-motion SpO2
  • SpO2% Range: 0 ~ 100%
  • SpO2 Accuracy: ±2% (70 ~ 100%,non-motion)
  • ±3% (70 ~ 100%, motion)
  • Pulse Rate Range: 30-250 bpm
  • Pulse Rate Accuracy: ±2 bpm(non-motion), ±3 bpm (motion)
  • Alarm Upper-lower Limit: Upper limit 70 ~ 100%, Lower limit 70 ~ 100%
  • SpO2 Probe: Red light LED wavelength: 660nm±5nm
  • Infrared light LED wavelength: 940nm±10nm
  • Standards: Meets performance standards of EN ISO 9919:2005

Rapid Temperature (Option)

  • Temperature Measurement Range: 30°C to 43°C (86°F to 109°F)
  • Typical Measurement Times
  • Oral (Quick Mode):3-5 seconds (non-fever temps), 8-10 seconds (fever temps):
  • (after insertion into measurement site):
    • Oral (Standard Mode): 6-10 seconds
    • Axillary Mode: 8-12 seconds
    • Rectal Mode: 10-14 seconds
    • Direct Mode (All Sites): 60-120 seconds
  • Approx. 6000 temperature readings
  • Standards: Meets performance standards of EN 12470-3:2000, ASTM E1112:2006
  • Pulse Timer: 60 Second count with a "beep" at 15 seconds, 2 "beeps" at 30 seconds, 1 "beep" at 45 seconds, and 2 "beeps" at 60 seconds
  • Patient Accuracy: A Standard Prediction Mode reading and a Direct Mode reading will differ by less than ±0.2°C (±0.4°F) on 98% of tested patients
  • Batteries:
    • Four "AA" Required.
    • Standard IEC package size.
    • Alkaline --1.5 Volt

EtCO2 (Option)

  • Mode of Sampling: Sidestream or Mainstream
  • Principle of Operation: Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts.
  • CO2 measurement Range: 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
  • CO2 Calculation Method: BTPS (Body Temperature Pressure Saturated)
  • CO2 Resolution: 0.1mmHg (0-69mmHg), 0.25mmHg (70-150mmHg)
  • CO2 Accuracy:
    • 0 ~ 40 mmHg ± 2 mmHg
    • 41 ~ 70 mmHg ± 5% of reading
    • 71 ~ 100 mmHg ± 8% of reading
    • 101 ~ 150 mmHg ± 10% of reading
    • Above 80 breath per minute ± 12% of reading
  • Sampling rate: 100Hz
  • Respiration Rate: 2 ~ 150 bpm
  • Respiration Rate accuracy: ±1 breath
  • Response Time: <3 seconds -
  • includes transport time and rise time
  • Inspired CO2 measurement Range: 3 ~ 50 mmHg
  • Standards: Meets performance standards of ISO/FDIS 21647:2004 (E), ASTM F1456-01, IEC/CDV 60601-2-55

Networking

  • Wired Networking:
    • Industry standard 802.11b/g wired network
    • Frequency Range: 2.412 ~ 2.484 GHz
    • Connected bedside number: Up to 16 bedside monitors
  • Wireless Networking:
    • Up to 100m indoors
    • Industry standard 802.11b/g wireless
    • Supports TCP/IP and UDP/IP Protocols

Power

  • Source: External AC power or internal battery
  • AC Power: 100 ~ 240VAC, 50/60Hz, 150VA
  • Battery: Built-in and lithium Ion rechargeable, 12.6V/5Ah
  • Charge Time: 8 hours
  • Operating Time: 3 hours

Environmental

  • Temperature: Operating: 5 ~ 40 °C
  • Storage: -10 ~ 45 °C
  • Humidity Range: Operating: ≤80 %
  • Storage: ≤80 %

FUSE

  • 3.15A/250V

LCD

  • Display Type: TFT color LCD
  • Size (diagonal): 5.0 inch
  • Active Area: 152.4 (W) × 91.44 (H) mm
  • Color arrangement: RGB-stripe
  • Dot pitch: 0.0635(W) × 0.1905(H) mm
  • Display Mode: Normally white, Transmissive
  • Interface: Digital (TTL)
  • Surface Treatment: Anti-Glare

Touchscreen

  • Type: Four-Wire Analog Resistive Touch Panel
  • Input Mode: Stylus Pen or Finger
  • Connector: FPC
  • Insulation resistance: 25Mω
  • Voltage: 7VDC
  • Chattering: 10ms
  • Transparency: 80%
  • Surface hardness: 3H
  • Durability-surface scratching: Write 100,000
  • Active force: 80gf
  • Knock Test: 1,000,000 times

Accesorios y Suministros

Cánula nasal

View all

Cánula de Muestra de CO2 con Entrega de O2 Dividida

P04092D - Male Luer, P03082D Female Luer

Web ID: 453