EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Restaurado - Sarns 7000 Carretera de circunvalaci

Restaurado - Sarns 7000 Carretera de circunvalaci

Web ID: 218
Condición: Refurbished

0 Artículo está en su carrito Ver carrito ahora

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

El Sarns 7000 Carretera de circunvalación Pulmonar de Corazón está disponible restaurado de DRE.

Specifications

Dimensions (with stand)

  • Height: 46.75 inches (118.7 cm) maximum, 34.75 inches (88.3 cm) minimum
  • Width: 22.25 inches (56.5 cm)
  • Length: 40.25 inches (102 cm)

Power Requirements

  • AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
  • DC: 12 VDC, 115 Watts

Internal Battery

  • Type: gel type, sealed and rechargeable.
  • Voltage: 12 VDC nominal
  • Quantity: 2
  • Charging Time: two batteries take 20 minutes to recharge from fully discharged.
  • Life expectancy: minimum 200 full charge/discharge cycles.

Chassis Leakage Current

100µA or less

Alarms

  • High Temp: Actuates at displayed incubator temperature of 39° ± 0.5 °C.
  • Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
  • Power Failure: Actuates if main AC power fails and no DC back-up power is present.
  • Heater Temp: Actuates if heater temperature exceeds 77°C nominal.
  • Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.

Temperature

  • Set Point Range: 21.5°C ± 1.0°C to 38 °C
  • Rise Time: 30 minutes
  • Variability: 0.2°C
  • Overshoot: 1.5°C maximum
  • Uniformity: 1°C
  • Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1°C
  • Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5°C
  • Observation Light Output: 10-foot candles at mattress level

-------------------------------------------------

The Sarns 7000 Heart Lung Bypass is available refurbished from Avante.