EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Tecnolog

Marca: Generadores Valleylab Electrosurgical Renovados y de segunda
Web ID: 739
Condición: Refurbished
Número de producto: 70202FX-N - New; 70202FXC - Used

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Force FX-C
Tecnología Instant Response™
La tecnología Instant Response™ provee a los cirujanos una mejora en el desempeño de ajustes de energía más bajos, reduciendo al mínimo el riesgo de daño del tejido fino y del estímulo neuromuscular, y disminuyendo la necesidad de incrementar el poder del generador.Un corte más liso a través de todos los tipos del tejido fino.
El generador de la Unidad Force FX™ ajusta automáticamente, respondiendo al cambio de tejido fino, manteniendo el mismo nivel de energía, y reduciendo al mínimo la fricción.La seguridad y la confiabilidad agregadas reduce al mínimo los riesgos de Electro quirúrgicos.
El acoplador capacitivo es reducido por 30-50% al usar la tecnología Instant Response™. Esta reducción es alcanzada limitando el voltaje del RMS y los armónicos de la alta frecuencia. Menos voltaje significa menos estímulo neuromuscular y entrega más exacta de la energía y así reducir daño colateral.

Specifications

Output Waveforms

  • Bipolar
    • Precise: 470 kHz sinusoid
    • Standard: 470 kHz sinusoid
    • Macro: 470 kHz sinusoid
  • Monopolar Cut
    • Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
    • Pure: 390 kHz sinusoid
    • Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
  • Monopolar Coag
    • Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
    • Fulgurate: 390 kHz damped sinusoid bursts with a repetition frequency of 30 or 57 kHz into 500 ohms.
    • Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21 kHzz < f < 35 kHz. Output is further modulated by a random 250 Hz envelope with a variable duty cycle.

Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208-264 volts (at rated load).

Low Frequency Leakage (50-60 Hz)

Source current, patient leads, all outputs tied together.

  • Normal polarity, intact chassis ground: < 10 µA
  • Normal polarity, ground open: < 50 µA
  • Reverse polarity, ground open: < 50 µA
  • Sink current, 140V applied, all inputs: < 50 µA

PER = 98

PER (Power Efficiency Rating) is the measure of an electrosurgical generator's ability to accurately deliver the selected power into a wide range of tissue types.

Physical Weight & Dimensions

  • Height: 4-3/8 inches
  • Width: 14 inches
  • Depth: 17 inches

Input Power Requirements

Operating range is 85 to 132 AC volts. Maximum current is 7 amperes in Cut and 4 amperes in Coag.

High Frequency Leakage

  • Bipolar: Less than 60 mA (rms)
  • Monopolar: Less than 150 mA (rms)

REM™ Contact Quality Monitoring System

  • Measurement Frequency: 80 kHz ± 10 kHz
  • Measurement Current: Less than 10 µA
  • Acceptable Resistance Ranges:
    • REM™ pad – 5-135 ohms
    • Non-REM™ pad – less than 20 ohms
  • Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.
  • Adaptive REM™: REM™ trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.

Output Characteristics

Bipolar

  • Precise Mode:
    • Maximum P-P Voltage: 450
    • Rated Load (OHMS): 100
    • Maximum Power (Watts): 70
    • Crest Factor* (Typical): 1.5
  • Standard Mode:
    • Maximum P-P Voltage: 320
    • Rated Load (OHMS): 100
    • Maximum Power (Watts): 70
    • Crest Factor* (Typical): 1.5
  • Macro Mode:
    • Maximum P-P Voltage: 750
    • Rated Load (OHMS): 100
    • Maximum Power (Watts): 70
    • Crest Factor* (Typical): 1.5

Monopolar Cut

  • Low Mode:
    • Maximum P-P Voltage: 1350
    • Rated Load (OHMS): 300
    • Maximum Power (Watts): 300
    • Crest Factor* (Typical): 1.5
  • Pure Mode:
    • Maximum P-P Voltage: 2300
    • Rated Load (OHMS): 300
    • Maximum Power (Watts): 300
    • Crest Factor* (Typical): 1.5
  • Blend Mode:
    • Maximum P-P Voltage: 3300
    • Rated Load (OHMS): 300
    • Maximum Power (Watts): 200
    • Crest Factor* (Typical): 2.5

Monopolar Coag

  • Desiccate Mode:
    • Maximum P-P Voltage: 3500
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 5
  • Fulg. High Crest Factor* Mode:
    • Maximum P-P Voltage: 8500
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 7.0
  • Fulg. Low Crest Factor Mode:
    • Maximum P-P Voltage: 6900
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 5.5
  • Spray Mode:
    • Maximum P-P Voltage: 9000
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 8

Monopolar Coag

  • Desiccate Mode:
    • Maximum P-P Voltage: 3500
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 5
  • Fulg. High Crest Factor* Mode:
    • Maximum P-P Voltage: 8500
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 7.0
  • Fulg. Low Crest Factor Mode:
    • Maximum P-P Voltage: 6900
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 5.5
  • Spray Mode:
    • Maximum P-P Voltage: 9000
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 120
    • Crest Factor* (Typical): 8

CEM™ Mode

  • Monopolar Cut (Low):
    • Maximum P-P Voltage: 1000
    • Rated Load (OHMS): 300
    • Maximum Power (Watts): 100
    • Crest Factor* (Typical): 1.5
  • Monopolar Coag (Dessicate):
    • Maximum P-P Voltage: 3500
    • Rated Load (OHMS): 500
    • Maximum Power (Watts): 70
    • Crest Factor* (Typical): 5

* Crest Factor is an indicator of a waveform's ability to coagulate without cutting.

Accesorios y Suministros

Avante Disposable Monopolar Handswitching Pencil (50/Box)

P00507.50

Web ID: 2089

Reusable Monopolar Handswitching Pencil

P01147

Web ID: 12358

Disposable Foot Control Pencil w/Blade Electrode

P01090

Web ID: 12363

Sterile Holsters for ESU Pencils

P05305

Web ID: 12370

Disposable Scratch Pad for ESU Pencils

P02050

Web ID: 12371

Bipolar Foot Pedal

P99002

Web ID: 12630

Mobile ESU Cart

P01594

Web ID: 12631

Disposable Split Grounding Pads w/ Cord Attached

P00084

Web ID: 12633

Disposable Split Grounding Pads w/o Cord

P00985

Web ID: 12374

Disposable Solid Grounding Pad w/ Cord Attached

P00085

Web ID: 12634

Disposable Solid Grounding Pad w/o Cord

P00910

Web ID: 12375

Reusable Grounding Plate - Pediatric

P00087

Web ID: 12635

Reusable Grounding Plate Cord

P00892

Web ID: 12376

Grounding Plate Cord - Disposable Pads

P00894

Web ID: 12632

Conmed Reusable Plate Cord-Single Plug

P00088

Web ID: 2032

Coated 3-1/2" McPherson Straight Micro (.5mm)

P01967

Web ID: 12378

Smooth 8" Bayonet Medium (2.0mm)

P00891

Web ID: 12379

Smooth 7-1/4" Bayonet Fine (1.0mm)

P01071

Web ID: 12380

Bovie Smooth 7" Straight Fine (1.0mm)

P00963

Web ID: 12381

Smooth 4-3/4" Straight Medium (2.0mm)

P00890

Web ID: 12382

Smooth 4-1/2" Curved Micro (.5mm)

P01051

Web ID: 12383

Reusable Bipolar Forcep Cord

P02039

Web ID: 12636