EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Datex Ohmeda 3800 Pulse Oximeter

Datex Ohmeda 3800 Pulse Oximeter

Marca: Datex-Ohmeda Medical Equipment
Web ID: 995
Condición: Retired

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The 3800 Pulse Oximeter has an easy-to-use interface and intuitive feature set, as well as advanced TruTrak® + technology, a proprietary clinical motion algorithm.

Features

  • TruTrak+ technology — a proprietary clinical motion algorithm which identifies, quantifies and corrects clinical patient motion
  • Always-visible Relative Perfusion Index™ (PIr) — a quick numeric value indicating pulse strength
  • Plethysmographic waveform reinforces clinical decisions
  • Backlit, always-on screen illuminate large numbers for fast recognition of readings
  • Dedicated function keys and an always visible, simple menu system with no hidden functions
  • Continuously displayed alarm limits for decisions at-a-glance
  • Exceeds international durability standards through rigorous stress testing during design and manufacturing processes
  • Full compatibility with all patient types and clinical situations using OxyTip®+ sensors and cables

Specifications

Physical Weight & Dimensions

  • Height: 3.7 inches (9.4 cm)
  • Width: 9.5 inches (24.4 cm)
  • Depth: 8.9 inches (22.5 cm)
  • Weight: 6.1 lbs (2.8 kg)

SpO2

  • Calibration: Functional
  • Accuracy (1 standard deviation): 70 to 100% (±2 digits); 70 to 100% (±3 digits during conditions of clinical patient motion with TruTrak+ enabled); Below 70% (unspecified)
  • Range: 0 to 100%
  • Resolution: 1%

Pulse Rate

  • Range: 30 to 255 beats per minute (bpm)
  • Accuracy (assuming a constant pulse rate): ±2% or±2 bpm (whichever is greater)
  • Accuracy (during conditions of clinical patient motion): Unspecified
  • Resolution: 1 bpm

Relative Perfusion Index (PIr) pulsatile value

  • Range: 0.00 to 9.99
  • Averaging interval: 12 seconds
  • Resolution: 0.01

Alarm Limits

  • Low SpO2 = OFF or 50 to 100& (preset 85%)
  • High SpO2 = 50 to 100% or OFF (preset OFF)
  • Low pulse = OFF or 30 to 235 bpm (preset: adult = 40 bpm, neonate = 100 bpm)
  • High pulse = 30 to 235 bpm or OFF (preset: adult = 130 bpm, neonate = 200 bpm)

Alarm Silence

2 minute silence (automatic, 2 min. on start up), all mute capability.

Signal Validation

Relative Perfusion Index pulsatile value PerfTrak® waveform display.

Memory

12 hour data storage for SpO2, pulse rate and alarm conditions (1 data point every 6 seconds).

SpO2 Calculation

  • 30 values per second/60 Hz
  • 25 values per second/50 Hz

Display

  • Digital - LED
  • Waveform - Backlit LCD

SpO2 Averaging Modes

Selectable 3, 6 or 12 seconds (short, medium, long) preset = 12 seconds (long)

Diagnostics

Automatic self test on start up, continuous memory tests during operation

Outputs

Serial (SpO2, pulse rate, alarms, error messages and trend data)

Electrical Specifications

  • Battery: 8 volt, 3.2 ampere-hours sealed lead-acid. Continuous charge while operating on AC
  • Charge Time: 100% capacity ≈ 8 hours; 80% capacity≈ 4 hours
  • Operation Time: ≈ 5.5 hours on fully charged battery
  • Low Battery Indicator: "Low Battery" displayed when < 5 minutes remaining
  • Overload Protection: 2 ampere 250V fuses
  • Power Consumption: 15 watts typical
  • Input Voltage: 100 to 240 V, 50/60 Hz