EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Baxter Colleague Volumetric Infusion Pump

Baxter Colleague Volumetric Infusion Pump

Brand: Baxter
Web ID: 1235
Condition: Retired

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Baxter's Colleague Volumetric Infusion Pump is designed to meet the fluid delivery needs of today's evolving healthcare environment.

The Colleague 3 Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications.

Fluid delivery applications include:

  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.

The Colleague Volumetric Infusion Pump feature spectrum includes:

  • Micro and Macro Rate Range
  • Basic Delivery Programming
  • Piggybacking Secondary Medications
  • Special Programming functions for Dosing

Specifications

General

  • Ac power Requirements: 100-120 VAC, 60Hz
  • Leakage Current: Less than 300 u amps earth leakage (tested per UL 2601)
  • External Fuses: 1.6 amp Type T (time delay) 250V
  • Power Cord: 2.9 m (9 feet) long with integrally molded hospital grade plug
  • Battery Supply System: Colleague 3 Pump - 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries) 12 hours to recharge to 80% of capacity Internal charge system recharges batteries whenever the device is connected to an AC outlet Range of Programmable Flow Rates Primary Infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments 1 to 1200 mL/hr in 1 mL/hr increments Piggyback Infusion on each pump channel 0.1. to 99.9 mL/hr in 0.1 mL/hr increments 1 to 500 mL/hr in 1 mL/hr increments Note: Rate limits can be configured for values less Than those stated above.
  • Volume to be Infused: 0.1 - 99.9 mL in 0.1 mL increments 1- 9999 mL mL/hr in 1 mL increments. Note: Volume to be Infused limits can be configured for values less than those stated above
  • KVO: 0.1 - 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate whichever is less.
  • Priming rate: 500 mL/hr (to be used only when not connected to patient)
  • Advance Air Rate to Advanced an Air Bubbles: At the same rate programmed for the current primary or piggyback infusion.
  • Air Bubble Setting: Measured accumulations of approximately
  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.