Baxter's Colleague Volumetric Infusion Pump is designed to meet the fluid delivery needs of today's evolving healthcare environment.

The Colleague 3 Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications.

Fluid delivery applications include:

  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.

The Colleague Volumetric Infusion Pump feature spectrum includes:

  • Micro and Macro Rate Range
  • Basic Delivery Programming
  • Piggybacking Secondary Medications
  • Special Programming functions for Dosing

Specifications

General

  • Ac power Requirements: 100-120 VAC, 60Hz
  • Leakage Current: Less than 300 u amps earth leakage (tested per UL 2601)
  • External Fuses: 1.6 amp Type T (time delay) 250V
  • Power Cord: 2.9 m (9 feet) long with integrally molded hospital grade plug
  • Battery Supply System: Colleague 3 Pump - 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries) 12 hours to recharge to 80% of capacity Internal charge system recharges batteries whenever the device is connected to an AC outlet Range of Programmable Flow Rates Primary Infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments 1 to 1200 mL/hr in 1 mL/hr increments Piggyback Infusion on each pump channel 0.1. to 99.9 mL/hr in 0.1 mL/hr increments 1 to 500 mL/hr in 1 mL/hr increments Note: Rate limits can be configured for values less Than those stated above.
  • Volume to be Infused: 0.1 - 99.9 mL in 0.1 mL increments 1- 9999 mL mL/hr in 1 mL increments. Note: Volume to be Infused limits can be configured for values less than those stated above
  • KVO: 0.1 - 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate whichever is less.
  • Priming rate: 500 mL/hr (to be used only when not connected to patient)
  • Advance Air Rate to Advanced an Air Bubbles: At the same rate programmed for the current primary or piggyback infusion.
  • Air Bubble Setting: Measured accumulations of approximately
  • Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
  • Whole blood and blood products.