EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

DRE ASG-120 Electrosurgical Unit

DRE ASG-120 Electrosurgical Unit

Brand: Avante
Web ID: 208
Condition: Retired
Product Number: 71250ASG - New 71250ASGR - Fac Rfr

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

One of the major benefits of the DRE ASG-120 electrosurgical unit is that it delivers the absolute highest level of safety and quality, but does not waste money on features you rarely use. The savings are found in the output power of 120 watts. Nearly every hospital specification on an electrosurgical generator calls for 300 watts of power. Have you ever asked yourself why you need 300 watts at all stations when it is used so infrequently? We did. It results in spending a few thousand dollars more per unit for a feature that just is not used.

The ASG-120 unit is a multipurpose electrosurgical generator for use in the modern physician's office and surgi-center. It features both monopolar and bipolar functions to satisfy all of your surgical demands with safety, flexibility, reliability, and convenience.

The ASG-120 features Cut, Blend, Coagulation, Fulguration, and Bipolar modes. It incorporates automatic safety features like self-test circuits, audible tones, discreet outputs, isolated circuitry, and Bovie NEM (neutral electrode monitoring).

Features

  • 120 watts maximum power output
  • Monopolar and bipolar capable
  • Coagulation and fulguration modes
  • Pinpoint and spray coagulation
  • Split or solid return electrodes
  • Delivers consistent, repeatable power into varying load impedances.
  • Defibrillator-proof type CF equipment
  • Footswitch indicator on face of unit
  • Dual voltage 50/60 Hz
  • Does not require calibration
  • Two-year manufacturer's warranty

Ready to Use

We wanted to make sure that when you take the DRE ASG-120 out of the box, it will be ready to go. It includes the hospital grade power cord, user's guide, one disposable electrosurgical pencil (standard push button), three electrodes, one reusable grounding cord, and five disposable split grounding pads.

Coagulation Modes

The DRE ASG-120 offers two modes of coagulation: Coagulation and Fulguration. Coagulation (pinpoint) provides precise control of bleeding in localized areas. Fulguration (spray) provides greater control of bleeding in highly vascular tissues over broad surface areas.

Designed for Safety

Electrosurgical safety is a combination of good equipment design and safe surgical practices. The ASG-120, as part of this combination, represents state-of-the-art design using the latest digital components. The ASG-120 also self-test circuits, audible tones, discreet outputs, and isolated output circuitry.

BovieNEM means safety is digitally designed inside. The FCFS (first come first served) discrete output design of the ASG-120 enhances safety by allowing only one output to be activated at any given time. This feature assures that only the device you first activate will be an active device. Secondary commands will not override the first command. As an example: while the monopolar foot controlled output is activated, all handswitching is inactive, as well as the bipolar footswitch capability. For additional safety this unit has totally separate bipolar controls, discrete output, dispersive electrode fault alarms in both sensing and non-sensing modes, and is designed with an isolated RF output.

Digital Error Detection

Digital error detection means unsurpassed safety for the surgeon, O.R. staff, and patient. The ASG-120 constantly monitors every aspect of it's output. At the sign of any problem the machine instantly disables the output and displays the appropriate error code on the display.

Built to the Required Standards

The ASG-120 has passed the following safety standards:

CSA C22.2 NO 601.1-M90, UL 2601-1-UL, IEC 60601-2-2 (1998-09), CENELEC EN 60601-1, CENELEC EN 60601-1-2, CENELEC EN 60601-2-2, FCC PART 15 Class A.

Optional Accessories at an Additional Charge

  • BiPolar Footswitch required for BiPolar operation.
  • Monopolar Footswitches
  • Pens / Pencils
  • Grounding Pads

Specifications

General

  • Classification: Class 1 Equipment, IEC 60601-1
  • Type: BF (Defibrillator Proof)
  • Spillage Protection: Drip Proof (IEC 60601-2-2)
  • Output Configuration: Isolated (RF Floating)
  • Cooling: Natural Convection, No Fan
  • Warranty: Two Years

Physical Weight & Dimensions

  • Height: 6 inches (15.2 cm)
  • Width: 10.25 inches (26 cm)
  • Depth: 12 inches (30.5 cm)
  • Weight: < 14 lbs (< 6.5 kg)

Power

  • Cut: 120 W @ 300 Ω
  • Blend: 90 W @ 800 Ω
  • Pinpoint: 80 W @ 1000 Ω
  • Spray: 40 W @ 1000 Ω
  • Bipolar: 30 W @ 200 Ω
  • Duty Cycle: 10 S / 30 S
  • Output Frequency: 490 kHz ± 5
  • Input Current: 4.5 A ~
  • Line Voltage: 100 - 240 V ~
  • Line Frequency: 50-60 Hz

Accessories & Supplies

Monopolar Foot Pedal

71253

Web ID: 12372

Bipolar Foot Pedal

P99007

Web ID: 12373

Mobile ESU Cart

7JMC09

Web ID: 12384

Disposable Split Grounding Pads w/ Cord Attached

P00084

Web ID: 12633

Disposable Split Grounding Pads w/o Cord

P00985

Web ID: 12374

Disposable Solid Grounding Pad w/ Cord Attached

P00085

Web ID: 12634

Disposable Solid Grounding Pad w/o Cord

P00910

Web ID: 12375

Reusable Grounding Plate - Pediatric

P00087

Web ID: 12635

Reusable Grounding Plate Cord

P00892

Web ID: 12376

Grounding Plate Cord - Disposable Pads

P00894

Web ID: 12632

Reusable Grounding Plate - Adult

P00088

Web ID: 12655