Invivo MDE Prism

Invivo MDE Prism

Brand: Invivo
Type: EKG Machines
Web ID: 1214
Condition: Retired
Product Number: UQMS1000

Need help?

Get an Equipment Quote

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

(800) 449-5328

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

MDE Escort Prism SE model 20414-103 patient vital signs monitor.

MDEEscort Prism SE model 20414-103 vital signs patient monitor with ecg/resp, Nellcor SPO2, NIBP, temp, turbo-temp, & printer; 900 MHz transceiver available.

Large color screen. The portable MDE Escort Prism SE has a battery and is easily transportable for any situation.This Escort Prism SE is the newer style of the popular Escort family, with BUILT-IN parameters including ecg/resp, SPO2, NIBP, temp, as well as Turbo-temp and a printer. The large color monitor screen is great and can be adjusted for optimal viewing. This unit is also available with the 900 MHz transceiver, model 20011V, which is to be used with the Vision Central Station system.

Display Type: CRT non-fade P31. Size: 6" (15.2 cm) diagonal. Number of Traces: 3. Trace Length: 3.2 sec @ 25 mm/sec

Front Panel Key Functions: Freeze, Record, Alarm Suspend, Alarm Set Up, Page Home, Next Page, 8 Softkeys, On/Standby.

Specifications

ECG

  • Range: 10 to 300 BPM
  • Accuracy: ± 2% or ± 2 BPM
  • Sweep Speed: 12.5, 25, 50 mm/sec selectable
  • High Alarm Limit Range: 50 to 250 BPM in 5 BPM steps – ADULT. 60 to 255 BPM in 5 BPM steps – PED/NEO.
  • Low Alarm Limit Range: 20 to 150 BPM in 5 BPM steps – ADULT. 20 to 150 BPM in 5 BPM steps – PED/NEO.
  • Bandwidth: 0.5 to 40 Hz (monitor); 0.05 to 100 Hz (diagnostic)
  • Leads: 3 lead: I, II, or III selectable. 5 lead, dual vector: I, II, or III, and V selectable
  • Sensitivity: 0.25 to 3 mV/cm selectable (12 levels)
  • Pacer Rejection: 0.1 to 2 msec; 2 to 700 mV
  • Heart Rate Source: ECG, Pleth, IBP
  • Defibrillation Tolerance: 400 joule with 50 ohm series lead impedance
  • Connector: Standard 6 pin MS
  • Isolation: Full electrical isolation

Respiration

  • Range: 4 to 200 BPM
  • Accuracy: ± 2% or ± 2 BPM
  • Lead: RA to LA
  • Sensitivity: 0.25 to 3 ohm/cm selectable
  • High Alarm Limit Range: 30 to 150, OFF – Adult. 50 to 200, OFF – PED/NEO
  • Low Alarm Limit Range: 5 to 50, OFF – Adult. 5 to 80, OFF – PED/NEO
  • Apnea Delay Alarms: 10, 15, 20, 30, 40 seconds, and OFF – ADULT. 10, 15, 20 seconds, and OFF – PED/NEO
  • Sweep Speed: 1.56, 6.25, 12.5, 25 mm/sec selectable
  • Respiration Rate Source: ETCO2 or Leads
  • CVA Rejection: Rejects most cardiovascular coincidence.
{