Details

THIS ITEM IS CURRENTLY UNAVAILABLE.  However for hospital use we do have newer refurbished options so please reach out to our team.  We also carry infant incubators dedicated for educational and simulation purposes only.

---

Integrates microprocessor-based controls with a modular design for efficient patient care

The GE Care Plus 4000 is an advanced incubator that provides superior thermoregulation with an integrated microprocessor-based controller. Optional active double-wall construction creates a precise and consistent environment for the neonate.

Features

  • Microprocessor-based control system for accurate thermoregulation
  • Self-test functions at power-up and during normal operation
  • Patient control (Servo) mode and air control (Manual) modes
  • Continuous positioning from 0° to 12°
  • Optional active double-wall construction

Specifications

Dimensions

  • Incubator with Elevating Base
  • Height: 54 - 62 in (136.2 - 157.5 cm)
  • Elevating base: 35 - 43 in (88.9 - 109.2 cm) mattress to floor
  • Depth: 28 in (71.1 cm)
  • Width: 33 in (83.8 cm)
  • Weight: 195 lbs (89 kg)
  • Incubator with Stationary Pedestal
  • Height: 53 in (135.9 cm)
  • Depth: 28 in (71.1 cm)
  • Width: 33 in (83.8 cm)
  • Weight: 180 lbs (80 kg)
  • Drawer Packages for Elevating Base or Stationary Pedestal
  • Overall: 11 x 11.5 x 18.25 in (28 x 29 x 46 cm)
  • Single deep drawer: 10 x 9 x 16.5 in (25 x 23 x 42 cm)
  • Dual drawer: 4.5 x 9 x 16.5 in (11 x 23 x 42 cm)
  • Incubator with Cabinet
  • Height: 55 in (141 cm)
  • Depth: 25.5 in (64.6 cm)
  • Width: 35 in (88.9 cm)
  • Weight: 185 lbs (84 kg)
  • Incubator Characteristics
  • Casters: 5 inch diameter, 2 locking, 2 non-locking
  • Mattress: 13.7 x 25.6 in (34.8 x 65 cm)
  • Mattress to Hood: 16 in (41 cm)
  • Door Height: 11 in (28 cm)
  • Positions: Continuous from 0° to 12°
  • Port holes: 6
  • Tubing ports: 6

Electrical

  • 120 VAC 60 Hz
  • Models: 115 V ± 10% 5.7 amps
  • Conforms to IEC 601.1, UL 544 and CSA 22.2 No. 125 requirements
  • Nominal power consumption: 450 watts at maximum heater output
  • Line voltage compensation: Input voltage is monitored and the heat output is adjusted to compensate for line voltage variations
  • Leakage currents: Less than 100 micro amps on 120V units
  • Leakage at patient probe: Less than 50 micro amps on 120V units

Environment

  • Operating temperature range: 68° - 86°F (20° - 30°C)
  • Storage temperature range: 13° - 140°F (-25° - 60°C)
  • Operating humidity range: 0 to 90%

System Controls

  • Microprocessor-based control system: Self-test functions are performed at power-up and during normal operation
  • Patient Control (Servo) Mode: 95° - 98.6°F (35° - 37°C)
  • Air Control (Manual) Mode: 73.4° - 98.6°F (20° - 37°C)
  • Up to 102.2°F (39°C) with control panel override switch
  • Air Velocity: Average < 10 cm/sec with inner wall
  • Temperature variability: Less than ± 0.2°F (0.1°C)
  • Resolution: ± 0°F (0.1°C)
  • Probe or interchangeability: ± 0.2°F (0.1°C)
  • Average Oxygen Input Concentration Range: 5 L/min 25-45% | 10 L/min 35-65% | 15 L/min 45 to 70%
  • Humidification: Standard — 10 to 75% dependent on nursery environment and incubator temperature setting | Servo — 40 to 85% regardless of nursery environment and incubator temperature setting