EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Valleylab Force 1C

Valleylab Force 1C

Brand: Valleylab
Web ID: 951
Condition: Retired

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Features

The Force 1C generator features a compact design suitable for the requirements for outpatient and office-based surgery, and the performance requirements for a full range of in-patient general surgery applications.

Digital Electrosurgical Unit suitable for all uses:

  • Outpatient procedures to full range surgical applications.
  • Monopolar and Bipolar outputs.
  • Audible and visual alarms.

Specifications

Physical Weight & Dimensions

  • Dimensions: 6 × 11 × 17 inches (15 × 28 × 43 cm)
  • 5 lbs (6.8 kg)

Output Waveform

  • Cut: 516 kHz sinusoid;
  • Blend 1: 516 kHz of sinusoid at 70% duty cycle recurring at 31 kHz;
  • Blend 2: 516 kHz bursts of sinusoid at 45% duty cycle recurring at 31 kHz;
  • Blend 3: 516 kHz bursts of sinusoid at 20% duty cycle recurring at 31 kHz;
  • Coag: 516 kHz damped sinusoidal bursts with a repetition frequency of 31 kHz;
  • Micropolar: 516 kHz sinusoid, unmodulated.

Output Characteristic

  • Cut: 2400 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 200 Nominal Power at rated load watts; 1.8 @ 100 W Crest Factor at rated load typical
  • Blend 1: 2800 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 175 Nominal Power at rated load watts; 2.7 @ 100 W Crest Factor at rated load typical
  • Blend 2: 3200 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 150 Nominal Power at rated load watts; 3.3 @ 100 W Crest Factor at rated load typical
  • Blend 3: 3600 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 125 Nominal Power at rated load watts; 4.6 @ 100 W Crest Factor at rated load typical
  • Coag: 5200 Max. (open circuit) P-P voltage; 300 Rated Load ohms; 75 Nominal Power at rated load watts; 8.0 @ 50 W Crest Factor at rated load typical
  • Micropolar: 900 Max. (open circuit) P-P voltage; 100 Rated Load ohms; 50 Nominal Power at rated load watts; 1.8 @ 40 W Crest Factor at rated load typical

Input Power Source

(Max. operating range: 85-135 or 170-270 VAC; Nominal operating range: 110-120 or 220-240 VAC)

  • Idle: 0.5 A Max. Current; 60 W Max. Power
  • Cut: 5.0 A Max. Current; 600 W Max. Power
  • Coag: 3.0 A Max. Current; 400 W Max. Power
  • Micropolar: 2.0 A Max. Current; 180 W Max. Power

Line Regulation

Between 85-135 volts or 170-270 volts input, output power into nominal load will vary no more than 20% or 5 watt, whichever is greater.

REM Contact Quality Mon.

Measurement frequency: 140 kHz ± 20 kHz. Measurement current: 3.0mA, maximum.

Low Frequency Leakage

Source current nominal polarity, intact chassis ground: <10 µA, ground open: <50 µA. Source current reverse polarity, ground open: <50 µA. sink current, 135 VAC applied, all inputs: <20 µ A. Chassis source current, ground open: <100 µA

Low Frequency Leakage

Micropolar RF leakage current: <150 mA. Monopolar RF leakage current: <150 mA