EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Respironics Lifecare PLV-100 Ventilator

Respironics Lifecare PLV-100 Ventilator

Marca: Respironics Equipo Respiratorio
Web ID: 1141
Condición: Retired

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Specifically designed for the long-term home use, the Respironics Lifecare PLV-100 is a reliable, easy to use portable volume ventilator. The PLV-100 is a versatile device that's effective on a wide range of patients including pediatric patients and patients with minute ventilation demands.

Features

Wide Range Of Applications
  • Use in home, sub-acute care andchronic care settings for transport
  • Small size and light weight ideal for mobile patients in wheelchairs, cars, and RVs
  • Capable of ventilating pediatric to adult patients
  • O2 bleed-in up to 40%
  • Provides volume or pressure ventilation
Digital Displays
  • Tidal Volume
  • I:E Ratio
  • Respiratory Rate
  • Inspiratory Flow Rate
Alarm Interface Capabilities
  • Remote Alarm
  • Pressure Alarm
  • Hospital/Nurse Call System
Three Modes Of Operation
  • Control l Assist-Control
  • Spontaneous Intermittent Ventilation (SIMV)
Multiple Power Sources
  • 120 VAC/240 VAC
  • 12 VDC external capabilities (up to 20 hours)
  • Internal battery (1 hour fully charged)

Specifications

Front Panel Controls

  • Mode Selection: Control, Assist/Control, SIMV
  • Tidal Volume: 0.05 to 3.00 liters ± 10%
  • Breath Rate: 2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2
  • Inspiratory Flow Rate: 10 to 120 L/Min
  • Battery Test Switch: Test Internal/External Battery
  • Main Power Switch: On/Off, Recharge
  • DC Circuit Breaker: Push/Pull Actuation
  • AC Circuit Breaker: Push/Pull Actuation
  • Airway Pressure Limit: 5 to 100 cm H2O ± 5 cm H2O
  • Patient Airway Pressure: -10 to 100 cm H2O ± 5
  • Sensitivity: +3 cm H2O ± 1 to at least -6 cm H2O
  • Low Pressure Alarm: 2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5

Alarms

The audible alarm ranges between 70 and 85decibles (at 1 meter) and has its own power sourceseparate from the internal battery. The power source and batteries are constantly recharged with a 2 amp trickle charge when the unit is plugged into AC power, whether it is operating or not.

  • Low Pressure: Visual/Continuous Audible
  • High Pressure: Intermittent Audible
  • Apnea: Visual/Continuous Audible
  • Inverse I:E Ratio: Visual Warning
  • IncreasedInspiratory Flow: Visual Warning
  • Low Internal Battery: Visual/Continuous Audible
  • Low External Battery: Visual/Continuous Audible
  • Reverse Battery Connection: Continuous Audible
  • Power Failure: Continuous Audible
  • Power Source: Audible
  • Unit Malfunction: Intermittent Audible