EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Bair Hugger 775 Patient Warmer

Bair Hugger 775 Patient Warmer

Marca: Bair Hugger
Web ID: 11894
Condición: Refurbished

0 Artículo está en su carrito Ver carrito ahora

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The Bair Hugger 775 Patient Warmer features four temperature settings, along with three temperature sensors.This patient warmer provides industry-leading forced-air warming in a surgical center or hospital. The Bair Hugger 775 Patient Warmer provides clinical warmth to the patient before, during and after surgery.

For more information about the Bair Hugger 775 Patient Warmer, contact an Avante representative today.

Features

  • Four temperature settings
  • Three temperature sensors and adjustable airflow provide safe, quick temperature response.
  • Hose-end temperature sensing ensures accurate air temperature delivery
  • Over-temperature, calibration, built-in hour meter and fault code reporting via the front panel no need to open the unit
  • Temperature in range indicator light
  • Fault indicator light
  • Over-temperature indicator light
  • Two preset fan speed settings allow for accurate airflow

Specifications

Dimensions

  • 13"h × 14"w × 13"d (33h × 35w × 33d cm)
  • Weight: 16lbs.

Operating Temperature

  • Operating temperatures: High: 43° ± 1.5°C (109.4° ± 2.7°F) Med: 38° ± 1.5°C (100.4° ± 2.7°F) Low: 32° ± 1.5°C (89.6° ± 2.7°F)

General

  • Leakage Current: Meets UL 60601-1 and IEC 60601-1 requirements. CAN/CSA-C22.2, No. 601.1
  • Device Ratings: 110-120 VAC, 50/60 Hz, 11.7 Amperes 220-240 VAC, 50/60 Hz, 7.2 Amperes 100 VAC, 50/60 Hz, 14.5 Amperes