EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

CASMED 740 Vital Signs Monitor

CASMED 740 Vital Signs Monitor

Marca: CASMED
Web ID: 1191
Condición: Refurbished

Need help?

Get an Equipment Quote

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

(800) 449-5328

Términos y Condiciones del Listado

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The CASMED® 740 monitor offers leading-edge technology in a rugged, portable, lightweight unit that weighs only 1.4 kg (3 lbs).

Easy to use, reliable, accurate, with multi-parameter capabilities

  • Fast, accurate blood pressure readings for neonates through adults . . . reduces the problems of motion in blood pressure measurement
  • Monitor oximetry and temperature either continuously or in a "spot-check" mode
  • Ideal for medical/surgical patients on pain medications including PCA pumps, endoscopy procedures, and for the EMS market

Featured Technologies

  • CASMED's proprietary MAXNIBP® technology

Oximetry choices include:

  • Masimo SET®
  • Nellcor® OxiMax®
  • Nonin®

Four models available:

  • CAS 740-1: MAXNIBP
  • CAS 740-2: MAXNIBP and SpO2
  • CAS 740-2T: MAXNIBP and Temperature
  • CAS 740-3: MAXNIBP, SpO2 and Temperature

Features

  • Created with emergency medical responders in mind
  • Wireless (infrared) printer option
  • User removable NiMH battery pack, internal power supply and DC power option
  • Flexible mounting solutions include cart, IV pole, bedrail, or wall; it can be fixed to a swivel wall mount designed specifically for emergency vechicles or carried in a protective, durable case

Optional Accessories

  • Infrared printer
  • RS-232 interface
  • Swiveled hard mount (for ambulance applications)
  • Roll stand and basket
  • Protective carrying case

Specifications

Physical Weight & Dimensions

  • Height: 6.75 inches (17 cm)
  • Width: 8.5 inches (21.5 cm)
  • Depth: 3 inches (7.5 cm)
  • Weight: 3 kg (1.4 lbs)

NIBP

  • Technique: Oscillometric (MAXNIBP® Technology)
  • Patient Range: Adult – Neonate
  • Measurement Range:
    • Systolic: 30-255 mmHg
    • Diastolic: 15-220 mmHg
    • MAP: 20-235 mmHg
  • Pulse Rate Range: 30-240 BPM
  • Accuracy:
    • Blood Pressure: ± 5 mmHg standard deviation no greater than 8 mmHg
    • Pulse Rate: ± 2% or ±2 BPM, whichever is greater

Oximetry (Options)

  • Masimo SET®
    • SpO2 Range: 0-100%
    • SpO2 Accuracy:
      • 70-100%, ± 2 digits (1 S.D.) Adult
      • 70-100%, ± 3 digits (1 S.D.) Neonate
    • Pulse Rate Range: 25-240 BPM
    • Pulse Rate Accuracy: ± 3 BPM
  • Nonin®
    • SpO2 Range: 0-100%
    • SpO2 Accuracy:
      • 70-100%, ± 2 digits (1 S.D.) Adult
      • 70-100%, ± 3 digits (1 S.D.) Neonate
    • Pulse Rate Range: 18-240 BPM
    • Pulse Rate Accuracy: ± 3% or ±1 digit, whichever is greater
  • Nellcor® OxiMax®
    • SpO2 Range: 1-100%
    • SpO2 Accuracy:
      • 70-100%, ± 2 digits (1 S.D.) Adult
      • 70-100%, ± 3 digits (1 S.D.) Neonate
  • Pulse Rate Range: 20-240 BPM
  • Pulse Rate Accuracy: ± 3 digits

Temperature (Optional)

  • Welch Allyn SureTemp® Plus:
    • Temperature Range: 29 °C to 42 °C (84.2 °F to 107.6 °F)
    • Accuracy: ± 0.1 °C (± 0.2 °F)
    • Meets or exceeds ASTM Standards

Control Panel

  • Display: LED display of measurement results, instructions, troubleshooting messages, and signal strength bar
  • Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate, %SpO2, Temperature (°C or °F)

Patient Alarms

Adjustable high and low alarms for Pulse Rate, SpO2, Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure

Operating Modes

  • History: Review of previous measurements
  • NIBP: Manual, STAT or Automatic (at preset intervals)
  • %SpO2: Continuous Monitoring
  • Temperature: Predictive or Continuous

Power

  • Source: External line or internal battery
  • AC Power: 100-240 VAC, 50-60 Hz, 0.5A; T1.25A L250V fuse rating
  • DC Power (EMS Option): 12-30 VDC
  • Battery: NiMH battery pack (removable)
  • Charge Time: 4 hours
  • Operating Time: Minimum 4 hours of continuous operation
  • Leakage Current: 100 _A (maximum)

Safety Limits

  • 145 mmHg (Neonate)
  • If measurement time exceeds 120 seconds (Adults), 90 seconds (Neonate)
  • If safety timer detects microprocessor failure

Operating Environment

  • Operating Temperature: 0°C to 50°C (32°F to 122°F)
  • Storage Temperature: -20°C to 65°C (-4°F to 149°F)
  • Humidity: 15 - 95%, non-condensing
  • Altitude: Up to 3048 meters (10,000 feet) if stored or used outside temperature and humidity ranges.

Standards

  • ANSI/AAMI SP-10
  • EN 60601-1, EN 60601-1-2, EN 865, EN 60601-2-30, EN 60601-2-49
  • ETL Listed – UL 60601, CAN/CSA C22.2 No. 601.1
  • CE Marking according to Directive 93/42/EEC

Accesorios y Suministros

Compatible ECG Cables, Leads, & Electrodes

3-Lead ECG Trunk Cable

Web ID: 11812