Specially Optimized for In-Hospital Transportation and Emergency Care 

The Avante Waveline EZ Plus patient monitor provides the perfect transportation solution. The multifunctional all-in-one parameter module and new application platform greatly enhance the efficiency of full clinical parameter monitoring for patient transfer. 

The Waveline EZ Plus is available through Avante Health Solutions. Contact an Avante representative today for more information. 

Features

  • 8.4" color TFT display with maximum 11 waveforms
  • Pacemaker detection, electrosurgical interference proof
  • Defibrillation protection and defibrillation synchronization
  • Optional, full touchscreen enables intuitive operation by clicking on the specific parameters or the waveform in real time, standard configuration includes the navigation knob
  • OxyCRG available to judge the respiration and circulation function for neonates
  • Large font display
  • Arrhythmia and S-T segment analysis
  • Built-in rechargeable Li-ion battery
  • Nurse Call function and bi-directional communication with MFM-CMS central station
  • SD card slot enables memory extension for large data storage
  • USB and serial ports enable abundant future upgrades
  • VGA output and analog output
  • Complete and flexible mounting solution for different scenarios 

Standard parameters: 3/5-lead ECG, Standard SpO2, RESP, NIBP, 2-Temp, PR 

Optional: Touchscreen, Thermal Recorder, Nellcor-SpO2, CO2 (sidestream & mainstream), 2-IBP, WLAN accessory, 12V DC input

Specifications

Dimensions

  • Length: 10.28” (261 mm)
  • Width: 7.8” (198 mm)
  • Height: 8.46” (215 mm)
  • Weight: 7.94 lbs. (3.6 kg) without battery 

Display

  • Color TFT LCD 8.4 Touchscreen Optional
  • Resolution: 800 X 600 dpi
  • Traces Displayed: up to 8
  • Waveforms Displayed: up to 11
  • Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s 
  • Various Working Interface Selectable
    • Standard Monitoring Display
    • Large Font Intensive Care Display
    • Trend Graph/ Monitoring Co-Display
    • Bed to Bed view Display (Optional)
    • OxyCRG Dynamic View Display
    • Drug Dose Calculation Interface 

RESP

  • Method: Trans-thoracic impedance
  • Operation mode: Auto/ Manual
  • RR Measurement range, Adult: 0 - 120 rPM
  • RR Measurement range, Neonate/Pediatric: 0 - 150 rPM
  • Resolution: 1rPM
  • Apnea alarm threshold: 10s, 15s, 20s (default), 25s, 30s, 35s, 40s
  • Alarm: 3 levels of audible and visual alarm, alarm events recallable
  • Band width: 0.2-2.5Hz (-3dB)
  • Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s 

ECG

  • Lead type: 5-lead and 3-lead selectable
  • 3 leadwire cable: RA, LA, LL or R, L, F
  • 5 leadwire cable: RA, LA, RL, LL, V or R, L, N, F, C
  • Lead selection: 3-lead: I, II, III,  
  • Lead selection, 5-lead: I, II, III, aVR, aVL, aVF, V
  • Gain selection: x0.125, x0.25, x0.5, x1, x2, auto
  • Sweep speed: 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
  • ECG HR Range, Adult/Pediatric: 15-300bpm
  • ECG HR Range, Neonate: 15-350bpm
  • Resolution & accuracy: ±1bpm or ±1%, whichever is greater
  • Filter:
    • Diagnostic mode: 0.05 - 150Hz
    • Monitoring mode: 0.5 - 40 Hz
    • Surgical mode: 1 - 20Hz
  • Protection: Withstand 5000VAC/50Hz voltage in isolation against defibrillation and electrosurgical interference
  • ST-Segment Detection:
    • Measurement range: -2.0 mV - 2.0mV
    • Alarm range: -2.0 mV - 2.0mV
  • ST-Segment Arrhythmia analysis and categorization: Yes
  • ARR Alarm: 3 levels of audible and visual alarm, alarm events recallable
  • Pace maker detection: Yes, and 5 types abnormal status detectable 

NIBP

  • Method: Automatic Oscillometric
  • Operation modes: Manual/Automatic/Continuous
  • Auto measurement time interval: Adjustable 1/2/3/4/5/10/15/30/60/90/120/240/480 Minutes
  • Measurement unit: mmHg/kPa selectable
  • Measurement types: Systolic, Diastolic, Mean
  • Pressure range for Adults:  
    • Systolic: 40 - 270 mmHg
    • Diastolic: 10 - 215 mmHg
    • Mean: 20 - 235 mmHg
  • Pressure range for Pediatrics:  
    • Systolic: 40 - 200 mmHg
    • Diastolic: 10 - 150 mmHg
    • Mean: 20-165 mmHg
  • Pressure range for Neonates:  
    • Systolic: 40 - 135 mmHg
    • Diastolic: 10 - 100 mmHg
    • Mean: 20 - 110 mmHg
  • Leak test and pressure auto calibration: Yes
  • Over-pressure protection: Dual Safety protection
  • Resolution: 1mmHg
  • Accuracy: Max mean error ±5 mmHg, max standard deviation ±8 mmHg
  • PR from NIBP: Measurement 40 - 240 bpm
  • Resolution: 1 bpm
  • Accuracy: ±3bpm or ±3%, whichever is greater
  • Leak test and pressure auto calibration: Yes 

SpO2

  • Measurement & Alarm Range: 0 - 100%
  • Resolution: 1%
  • Accuracy: ±2% (70-100%, Adult/Pediatric), ±3% (70-100%, Neonate)
  • PR Measurement and Alarm Range: 30 - 300bpm
  • Resolution: 1 bpm
  • Accuracy: 3bpm
  • Refresh rate: 2s 

Temperature (2 Channels, 1 probe by default)

  • Measurement /Alarm range: 0 - 50 (32-122 F)
  • Probe type: YSI (B Series) and CF-FI
  • Resolution: 0.1
  • Accuracy: ±0.1°C (without probe)
  • Channel: Dual-channel. Provide T1; T2; T