Details

Datex-Ohmeda 5330's are suitable for performing anesthetic vaporizer concentrations checks in the field. These units are available either used or professionally refurbished from DRE.

Features

  • Monitors 5 Agents: Enflurane, Halothane, Isoflurane, Sevoflurane, Desflurane
  • Easy-to-read digital display screen 5330

Specifications

Dimensions / Power

  • Height: 4 inches (10.16 cm)
  • Width: 10 inches (25.4 cm)
  • Depth: 16 inches (40.64 cm)
  • Weight: 13.25 lbs (6 kg)
  • Power Input: 110, 120 VAC, 50/60 Hz

Operational Environment

  • Temperature: 15 to 35°C
  • Humidity: 0 to 95% RH, non-condensing

Storage Environment

  • Temperature: -5 to 50°C
  • Humidity: 0 to 95% RH, non-condensing

Display

  • Displayed Parameter: Agent inspired & expired concentration levels (%)
  • Trendgraph: Last 120 breaths, 20 min, 60 min, 120 min
  • Sample Flow Rate: 250 mL/min
  • Response Time: 600 msec
  • Warm-Up Time: 5 minutes to operation
  • Compensation: Temperature & barometric pressure
  • Analog Output: Anesthetic agent waveform
  • Digital Output: RS-232 compatible

Enflurane, Halothane, Isoflurane, Sevoflurane

  • Range: 0 to 9%
  • Accuracy: 0 to 1.0% (± 0.2%)
  • 1.1 to 6.0% (± 0.3%)
  • 6.1 to 7.0% (± 0.35%)
  • Resolution: 0.1%
  • Alarms: High expired 0.2 to 10% agent concentration
  • High inspired 0.2 to 10% agent concentration
  • Low expired 0 to 9.8% agent concentration
  • Low inspired 0 to 9.8% agent concentration

Desflurane

  • Range: 0 to 24%
  • Accuracy: 0 to 4.9% (± 0.2%)
  • 5.0 to 9.9% (± 0.5%)
  • 10.0 to 18.0% (± 1.0%)
  • 18.1 to 22.0% (± 2.0%)
  • Resolution: 0.1%
  • Alarms: High expired 0.2 to 10% agent concentration
  • High inspired 0.2 to 10% agent concentration
  • Low expired 0 to 9.8% agent concentration
  • Low inspired 0 to 9.8% agent concentration