EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

Avante Waveline Pro Touch-Screen Anesthesia Monitor

Avante Waveline Pro Touch-Screen Anesthesia Monitor

Brand: Avante
Web ID: 1077
Product Number: 66009B2x (x = additional suffix for varying configurations

Need help choosing?

Ask an Equipment Expert

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

(800) 449-5328

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

Superior performance with a high-resolution touch-screen

The Avante Waveline Pro displays as many as eight colorful waveforms on a crisp, high-resolution 15-inch touch-screen that optimizes the speed of patient care. The user can make quick screen adjustments, set default settings, alarm limits, and manage up to 72 hours of detailed patient data. Designed for a fast paced work environment, the Waveline Pro patient monitor offers an extremely simple and adaptable user interface. Patient information along with vital-signs settings can be quickly modified to meet the needs of a patient's changing condition, converting from a basic vital signs monitor, to a continuous bed side monitor, to an operating room monitor. Now with Masimo SET® Pulse Oximetry!

This patient monitor offers standard measurements of: non-invasive blood pressure, 3-lead or 5-lead ECG with arrhythmia detection, motion tolerant SpO2, Temperature, and Respiration rate. End-tidal CO2, Anesthetic Agent measurement, Cardiac Output and Invasive blood pressure are available options.

The Waveline Pro monitor offers several connective solutions to network multiple monitors and/or manage patient data on an electronic medical records platform or a HL7-based hospital information system. It offers Ethernet and RS-232 connections with an open source communication protocol, and it is HL7 compliant. The HL7 network protocol will allow for all patient information and vital sign trends to be transferred and stored on a hospital information system.

Capnography is outpacing Pulse Oximetry as the new standard of care in outpatient surgery procedures. Click to see the dramatic difference between CO2 and SPO2.

Features

  • New - Masimo SET® Pulse Oximetry — ask your Avante Representative for details
  • Optional Integrated Dräger Anesthesia Gas Bench expands measurement capabilities
  • Optional agent bench is compact and part of main unit. All in one monitor!
  • Large (15"), touch-screen with high-resolution flat panel display
  • Displays as many as eight waveforms
  • Multi-lead simultaneous ECG monitoring
  • Automatically set alarm limits
  • Graphical and tabular trending
  • Color-coded alarms
  • Patient data entry
  • Networking capability
  • Works with EMR and wireless telemetry systems
  • Large time stamp in upper right corner
  • Standby button
  • Volume and Sounds are adjustable and can be turned off completely
  • Option to print Numerical Data Only
  • Color of the waveforms can be changed
  • Battery backup (two removable batteries)
  • Available with a built-in thermal printer

Specifications

Performance

  • Display: 15-inch color touch-screen
  • Trace: 8 waveforms
  • Indicator:
    • Alarm indicator
    • Power indicator
    • QRS beep and alarm sound
  • Trend time: 1 - 72 hour
  • Recorder:
    • Built-in, thermal array, 3 channels
    • Record width: 48mm
    • Recorder paper: 50mm
    • Record speed: 25mm/s, 50mm/s

ECG

  • Input: 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
  • Gain Choice : x0.5, x1, x2, x4
  • Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
  • ECG Waveforms: 7 channels
  • Penetration Voltage: 4000VAC 50/60Hz
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
  • HR Display Range: 30 ~ 300bpm
  • Accuracy: ±1bpm or ±1%, whichever is greater
  • Alarm Limit Range Setting:
    • Upper limit 100 ~ 200bpm,
    • Uower limit 30 ~ 100bpm

RESP

  • Measure Method: RA-LL impedance
  • Range: 0 ~ 120 rpm
  • Accuracy: ±3 rpm
  • Alarm Limit Setting:
    • Upper limit 6 ~ 120 rpm,
    • Lower limit 3 ~ 120 rpm
  • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)

NIBP

  • Measuring Technology: automatic oscillating measurement
  • Cuff Inflating: <30s (0 ~ 300 mmHg, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto
  • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
  • Pulse Rate Range: 30 ~ 250 (bpm)
  • Measuring Range:
  • Adult/Pediatric Mode: SYS: 40 ~ 250 (mmHg)
  • DIA :15 ~ 200 (mmHg)
  • Neonatal Mode: SYS: 40 ~ 135 (mmHg)
  • DIA : 15 ~ 100 (mmHg)
  • Accuracy:
    • Maximum Mean error: ±5mmHg
    • Maximum Standard deviation: 8mmHg

TEMP

  • Range: 25 ~ 50 (°C)
  • Accuracy:
    • ± 0.2°C (25.0 ~ 34.9°C)
    • ± 0.1°C (35.0 ~ 39.9°C)
    • ± 0.2°C (40.0 ~ 44.9°C)
    • ± 0.3°C (45.0 ~ 50.0°C)
  • Display Resolution: 0.1°C
  • Alarm Limit Setting:
    • upper limit 0 ~ 50°C,
    • lower limit 0 ~ 50°C
  • Channel: 2 channels

Masimo SET SpO2

  • SpO2 Accuracy (non-motion):
  • Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
  • SpO2 Accuracy (motion):
  • Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
  • SpO2: ±2 %
  • PR: ±3 bpm
  • Modes:
  • Averaging mode: 2,4,8,10,12,14 and 16 s
  • Sensitivity: Normal, APOD and Maximum
  • PR Accuracy (non-motion):
  • Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
  • Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
  • PR Accuracy (motion):
  • Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
  • Measuring Range:
  • SpO2: 1~100 %
  • PR: 25~240 bpm
  • Perfusion: 0.02~20 %

IBP

  • Measurement Range: -50 ~ 300mmHg
  • Channel: 2 channels
  • Pressure Transducer: sensitivity, 5µV/V/mmHg
  • Impedance Range: 300 ~ 3000ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1mmHg
  • Accurancy: ±1mmHg or ±2%,whichever is greater
  • AlarmRange: -10 ~ 300mmHg

EtCO2 (Optional)

  • CO2 Measurement Range: 0 ~ 99mmHg
  • Accuracy:
    • ±2mmHg (0 ~ 38mmHg)
    • 39-99mmHg ±5% of reading +0.08% for every 1mmHg (above 38mmHg)
  • Sampling Rate: 50 ml/min
  • Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes.
  • Respiration Rate: 0 ~ 150 breaths/min
  • Mode: adult, neonate

Cardiac Output (Optional)

  • Measurement Method: Thermodilution Method
  • CO2 Measurement Range:
    • C.O.: 0.1 to 20 L/min
    • TB: 23 to 43
    • TI: 0 to 27
  • Resolution:
    • C.O.: ±5% or ±0.1 L/min, whichever is greater, as measured using electornically generated flow curves.
    • TB, TI: 0.1
  • Accuracy:
    • ±5 or ±0.1 L/min, whichever is greater, as measured using electronically generated flow curves.
    • TB, TI ±0.1 (without sensor)
  • Alarm Range: TB 23 to 43
  • Repeatability: C.O. ±2% or 0.1 L/min, whichever is greater, as measured using electronically generated flow curves.

Anesthetic Agents (Optional)

  • Method: Infrared absorption
  • Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
  • Measurement Range:
    • Halothane, Isoflurane: 0 - 8.5%
    • Enflurane, Sevoflurane: 0 - 10%
    • Desflurane: 0 - 20%
    • CO2: 0 - 10%
    • N2O: 0 - 100%
    • O2: 0 - 100%
  • Bias:
    • Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ±(0.15 Vol% + 15% rel.)
    • CO2: ±(0.5 Vol% + 12% rel.)
    • N2O: ± (2 Vol% + 8% rel.)
    • O2: ±3 Vol%

Networking

Industry standard 802.11b/g wireless network

Power

  • Source: External AC power or internal battery
  • AC Power: 100 - 240VAC, 50/60Hz, 150VA
  • Battery: Built-in and rechargeable lithium ion
  • Operating TIme: 3+ hours

Environmental

  • Temperature:
    • Operating: 5° - 40° C
    • Storage: -20° - 65° C
  • Humidity Range:
    • Operating: ≤;80 %
    • Storage: ≤80 %

Other Standard Feature

OxyCRG, drug dose calculation, cascading ECG, On screen NIPB trends (up to 250 readings), user set defaults, Arrhythmia detection, ST segment

Videos

Accessories & Supplies

3 Lead ECG Cable

P05022

Web ID: 12625

3 Lead ECG Cable With Detachable Leads

P05020

Web ID: 12602

5 Lead ECG Cable

P05023

Web ID: 12626

5 Lead ECG Cable With Detachable Leads

P05021

Web ID: 12603

Adult Snap Electrodes Disposable

P00842

Web ID: 12604

NIBP Hose

P02112

Web ID: 12611

Rolling Stand

6600RD

Web ID: 12457

Wall Mount

P02116

Web ID: 12607

Base Plate

66011MP

Web ID: 12608

Printer

6600P

Web ID: 12609

Power Cord

P00100

Web ID: 12610

IBP Cable

P02118B2

Web ID: 12605

Disposable IBP Transducer

P01858B2

Web ID: 12606

Skin Temperature

P05054

Web ID: 12615

Oral/Rectal Temperature

P05054O

Web ID: 12616

Sidestream LoFlo CO2 Module

P02211

Web ID: 12461

Sidestream CO2 Module Bracket

P02211B

Web ID: 12613

Mainstream CO2 Modules

P02212

Web ID: 12627

Sampling Line w/ Dehumidification Tube

P01933S

Web ID: 12460

Tee Adapter for CO2 Sampling (10/pk)

P02115

Web ID: 12614

Mainstream Adult/Pediatric Airway Adapter

P00300

Web ID: 12628

Mainstream Neonatal Airway Adapter

P00301

Web ID: 12629

Analog SpO2 Finger Probe

P0375AFSEXT

Web ID: 12596

Analog SpO2 Extension Cable

P0395EXT

Web ID: 12597

Analog SpO2 Y-Site Probe

P02270AY

Web ID: 12598