EU Cookie Consent

We use cookies on this website to provide statistics that help us give you the best experience. By continuing to use the site without changing settings, you are agreeing to our use of cookies. Read Privacy Policy

DRE Axis 4 Portable Vital Signs Monitor

DRE Axis 4 Portable Vital Signs Monitor

Brand: Avante
Web ID: 930
Condition: Retired

Need help?

Get an Equipment Quote

(800) 462-8195

Biomedical HTM Customers

Service, Repair & Parts

(800) 449-5328

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

DRE's Light, Portable Monitor Measures Temp, NIBP, ECG and SpO2

At less than 1.3 kg (2.87 lbs), the DRE Axis 4 provides vital signs monitoring that is mobile and accurate. It has a crisp, easy-to-read monographic LCD screen that displays waveforms for ECG leads I, II and III. The Axis 4 also displays data for SpO2, heart rate, pulse rate, SYS, DIA, mean, temp, time, battery status, alarm status, contrast, beep and alarm sound status, and trends.

Features

  • Mobile monitoring: Provides up to three hours of on-the-move battery operation (up to five hours of battery operation for ECG-only use)
  • Designed for desktop, IV pole mount and handheld use
  • Demo mode for staff training
  • Monitors adult, pediatric and neonatal patients
  • Features a tabular trend display

Specifications

Physical Weight & Dimensions

  • Height: 3.3 inches (85 mm)
  • Width: 9.6 inches (245 mm)
  • Depth: 3.3 inches (84 mm)
  • Weight: 2.87 lbs. (1.3 kg)

Display

  • Screen Type: Monographic LCD
  • Resolution: 240 x 64 pixels
  • Waveform Type: ECG leads I, II, III
  • Indicator: AC power or battery operated indicators
  • Display Data: SpO2 , heart rate, pulse rate, SYS, DIA, mean, temp, time, battery status, alarm status, contrast, beep and alarm sound status, trend

Trends

  • Memory: 15 day at 10 second intervals
  • Display: Tabular

Interface

S/W Upgrade: RS-232

AC Power

  • Input Voltage: 100 ~ 240VAC, 50/60Hz
  • Output Voltage: 12VDC, 3.5A

Battery

  • Type: Li-ion
  • Battery Run Time: 3 hours (Up to 5 hours for ECG only)

Environmental Temperature

  • Operating: 5° ~ 40°C (41° ~ 104°F)
  • Storage: -20° ~ 60°C (-4° ~ 140°F)

Humidity

  • Operating: 15 ~ 95% , non-condensing
  • Storage: 10 ~ 95% , non-condensing

Alarms

  • Audible and Visual Alarms: SpO2, pulse rate, heart rate, SYS, DIA, mean, temp, low battery, ECG lead off, SpO2 fi nger off, system error
  • Silence: 2 minutes
  • Volume: 0 (off ) ~ 7

ECG Performance

  • Leads: 3-lead I, II, III
  • Display Sensitivity: 1.0 cm/mV ±10%
  • Gain Selection: x0.25, x0.5, x1
  • Frequency Response: 0.5 ~ 40H
  • Sweep Speed: 25 mm/sec, 50 mm/sec
  • Pacemaker Detection: Yes
  • Heart Rate Range: 30 ~ 300 bpm
  • Heart Rate Accuracy: ± 3% or ± 3 bpm (whichever is greater)

Pulse Oximetry Performance

  • Measurement Range:
    • Sp02: 0 to 100%
    • Pulse Rate: 25 to 300 bpm
  • Accuracy:
    • Sp02: 70 to 100% ±2%, 0 to 69% - unspecified
    • Pulse Rate: 25 to 250 bpm ±3 bpm

Temperature

  • Measurement Range: 15° to 45° C (59° to 113° F)
  • Accuracy: ± 0.1 C (25° to 45° C); ± 0.2 C (15° to 24° C)
  • Probe Type: Compatible with YSI 400 series

NIBP Performance

  • Technique: Oscillometric
  • Measurement Range - Large Patient; SYS: 30 ~ 270 mmHg; DIA: 10 ~ 220 mmHg; Mean: 20 ~ 225 mmHg
  • Measurement Range - Medium Patient; SYS: 30 ~ 235 mmHg; DIA: 10 ~ 220 mmHg; Mean: 20 ~ 225 mmHg
  • Measurement Range - Small Patient; SYS: 30 ~ 135 mmHg; DIA: 10 ~ 110 mmHg; Mean: 20 ~ 125 mmHg
  • Cuff Pressure Range: 0 ~ 280 mmHg
  • Pressure Accuracy: ±2% or ±3 mmHg (whichever is greater)
  • Static: ±5 mmHg average error
  • Clinical: 8 mmHg standard deviation
  • Pulse Rate Measuring Range: 40 ~ 240 bpm
  • Inflation Time for Cuff: Less than 40 sec. (standard adult cuff )
  • Total Cycle Time: 20 to 45 seconds typical (dependent on heart rate and motion artifact)
  • Intervals for AUTO Measurement Time (min): 5, 10, 15, 30, 60, 90, 120
  • Overpressure Protection: Hardware and software double protections
  • Adult: 315±10 mmHg
  • Pediatric: 265±10 mmHg
  • Neonate: 155±10 mmHg